Technical Specification

AirSentric™️ WB55 AR10-B-102-A

Carbon Dioxide Monitor

An Indoor Air Quality monitor

NDIR CO₂ Sensor

Sensing Range​
0 to 5,000 ppm / 0.04 to 2 % volume CO₂
Accuracy​​
+/- 3 % of reading

Temperature Sensor

Temperature range​
0 to 50°C​
Accuracy​​
(at –10 to 85 °C) ± 0.3 °C (typical)​
Stability
< 0.01 °C/Y​r

Humidity Sensor

Humidity range​
5 to 95 % RH continuous​
Accuracy​​
(at 0 – 90 % RH) ± 5 % RH (max)​
Stability
< 0.25 % RH/Yr​

Size & Weight

Size
Size (LxWxD) – 4 in (100 mm) x 6 in (150 mm) x 1.8 in (45 mm)
Weight
320 g (enclosure weight without power adapter)​

Connectivity

Zigbee
GY36 Zigbee to Ethernet Gateway, 2.4 GHz, up to 30 m range​
Wi-Fi​
802.11 b/g (optional subject to MOQ)

Power

Power Supply​
100 - 240 V International Adapter, DC 12 V +/- 10 %, 420 mA​
Power Cable​
2 metres​
Power Consumption​
250 mA @ 12 V​
Power over Ethernet​
PoE option available​

Operating Temperature

Range
0°C to 50°C (32 F to 122 F)​

Calibration

Auto-calibration​
CO₂ NDIR self-calibration​
Calibration service​
Typically not required within 5 years, annual service option available​

Accessories

Mounting
Wall mount bracket supplied​
Elbow connector​
Extended elbow connector (2 metres) – optional

Software

User interface​
HEX cloud-based account​

Compliance / Certification

Accreditation​
RESET Air accredited
Compliance
CE, FCC, RoHS